Product Details
Eis Function: Include
Cyclic Voltammetry: Include
Galvanostatic Charge and Discharge: Include
  • CS310 Electrochemical Workstation
  • CS310 Electrochemical Workstation
  • CS310 Electrochemical Workstation
  • CS310 Electrochemical Workstation
  • CS310 Electrochemical Workstation
Find Similar Products

Basic Info.

Model NO.
CS310
Current Range
±2A
Potential Control Range
±10V
Eis Frequency
10μHz~1MHz
Transport Package
Standard Carton
Specification
36.5(W)X30.5 (D)X16 (H)
Trademark
Corrtest
Origin
Wuhan City, China
HS Code
9027809900
Production Capacity
1000 Sets/Year

Product Description

CS310 electrochemical workstation (potentiostat / galvanostat) contains a fast digital function generator, high-speed data acquisition circuitry, a potentiostat and a galvanostat. With high performance in stability and accuracy with advanced hardware and well-functioned software, it is a comprehensive research platform for corrosion, batteries, electrochemical analysis, sensor, life science and environmental chemistry etc.        
Applications:                
(1) Reaction mechanism of Electrosynthesis, electrodeposition (electroplating), anodic oxidation, etc.
(2) Electrochemical analysis and sensor; 
(3) New energy materials (Li-ion battery, solar cell, fuel cell, supercapacitors), advanced functional materials, photoelectronic materials;
(4) Corrosion study of metals in water, concrete and soil etc;
(5) Fast evaluation of corrosion inhibitor, water stabilizer, coating and cathodic protection efficiency.
HARDWARE FEATURES
Dual-channel high-speed 16bit/high-accuracy 24bit AD converter;
Built-in frequency response analyzer(FRA), frequency range: 10μHz~1MHz;
High-bandwidth high input impedance amplifier unit
Built-in FPGA DDS digital signals synthesizer
High-power Potentiostat/Galvanostat/ZRA;
Potential control range: ±10V; Compliance voltage: ±21V;
Current control range: ±2A ;
Potential resolution: 10μV; current resolution 1pA.
SOFTWARE FEATURES
CS studio software provides users a versatile smoothing/differential/ integration kit, which can complete the calculation of peak height, peak area and peak potential of CV curves.
CS studio also provides powerful non-linear fitting on Butler-Volmer equation of polarization curve. It can calculate Tafel slope, corrosion current density, limitation current, polarization resistance, corrosion rate. It can also calculate the power spectrum density, noise resistance and noise spectrum resistance based on the electrochemical noise measurements.
CS Studio software can achieve real time saving of the measuring data. The data can be automatically saved even in case of sudden power off.
CS studio kit has a built-in versatile timing policy for combined measurements, which can facilitate the automation of experiments and save time.
Battery analysis: charge & discharge efficiency, capacity, specific capacitance, charge & discharge energy etc.
EIS analysis: Bode, Nyquist, Mott-Schottky plot
 

TECHNICAL ADVANTAGES
1. Impedance (EIS)
CS310 potentiostat applies correlation integral algorithm and dual-channel over-sampling technique, and has strong anti-interference ability. It is suitable for EIS measurements of high-impedance system (>109Ω, such as coating, concrete etc.). It can also be used to obtain Mott-Schottky curve and differential capacitance curve. During test, the software can display real-time open circuit potential(OCP) without entering.
  
EIS of AA6063 Al alloy in Ce3+ containing 3% NaCl solution(left)
Polarization curve of Ti-based amorphous alloy & stainless steel in 3%NaCl solution(right)

2. Polarization curve
It can complete linear polarization curve and Tafel plot measurements. The user can set the anodic reversal current (passivation film breakdown current) of the cyclic polarization curve to determine material's pitting potential and protection potential and evaluate the its susceptibility to intergranular corrosion. The software employs non-linear fitting to analyze polarization curve, and can make fast evaluation of material's anti-corrosion ability and inhibitors.

3. Voltammetry
It can do the following electroanalysis methods: Linear Sweep Voltammetry(LSV), Cyclic Voltammetry(CV),  It integrates calculation of peak area, peak current and standard curve analysis.

LSV curve: mesoporous carbon material in 0.1M KOH
 


CV curves of PPy supercapacitor in 0.5 mol/L H2SO4
 
4. Electrochemical Noise
With high-resistance follower and zero-resistance ammeter, it measures the natural potential/current fluctuations in corrosion system. It can be used to study pitting corrosion, galvanic corrosion, crevice corrosion, and stress corrosion cracking etc. Through noise spectrum, we can evaluate the inducement, growth and death of metastable pitting and crack. Based on calculation of noise resistance and pitting index, it can complete localized corrosion monitoring.

Electrochemical noise of low-carbon steel in 0.05mol/L Cl-+0.1mol/L NaHCO3
5. Full floating measurement
CS310 workstation uses full-floating working electrode. It can be used for autoclave electrochemical measurements, on-line corrosion monitoring of metallic components under the ground (rebar in concrete, etc.)

6. User-defined methods
CS310 workstation supports user-defined combination measurements. The user can set cyclic  timing measurements of an electrochemical method or several methods.
 We are able to provide API functions and development examples, which facilitates some users' requirements for secondary development and self-defined measurements.
 
Specifications
Support 2-, 3- or 4-electrode system
Potential control range: ±10V Current control range: ±2A
Potential control accuracy: 0.1%×full range±1mV Current control accuracy: 0.1%×full range
Potential resolution: 10μV (>100Hz),3μV (<10Hz) Current sensitivity:1pA
Rise time: <1μS (<10mA), <10μS (<2A) Reference electrode input impedance:1012Ω||20pF
Current range: 2nA~2A, 10 ranges Compliance voltage: ±21V
Maximum current output: 2A CV and LSV scan rate : 0.001mV~10,000V/s
CA and CC pulse width: 0.0001~65,000s Current increment during scan: 1mA@1A/ms
Potential increment during scan: 0.076mV@1V/ms SWV frequency: 0.001~100 kHz
DPV and NPV pulse width: 0.0001~1000s AD data acquisition:16bit@1 MHz,20bit@1 kHz
DA Resolution:16bit, setup time:1μs Minimum potential increment in CV: 0.075mV
IMP frequency: 10μHz~1MHz Low-pass filters: Covering 8-decade
Potential and current range: Automatic
PC and O/S requirement
 Interface: USB 2.0 Operating System: 
Windows 2000/NT/XP/ win7/win8/win10
Weight / Measurements
Net weight: 6.5kg Gross weight: 10kg
Instrument size: 36.5 x 30.5 x16cm; After package: 52.5 x 37.5x 22.5cm
 
Electrochemical Impedance Spectroscopy(EIS)
Signal generator
Frequency range:10μHz~1MHz AC amplitude:1mV~2500mV
DC Bias:-10~+10V Output impedance:50Ω
Waveform: sine wave, triangular wave and square wave Wave distortion:<1%
Scanning mode: logarithmic/linear, increase/decrease
Signal analyzer
Integral time:minimum:10ms or the longest time of a cycle Maximum:106 cycles or 105s
Measurement delay:0~105s
DC offset compensation
Potential automatic compensation range:-10V~+10V Current compensation range:-1A~+1A
Bandwidth: 8-decade frequency range, automatic and manual setting

Software
Techniques included in CS310 potentiostat
Stable polarization
Open Circuit Potential (OCP), Potentiostatic (I-T measurement), Galvanostatic, Potentiodynamic (Tafel plot), Galvanodynamic
Transient polarization
Multi-Potential Steps, Multi-Current Steps, Potential Stair-Step (VSTEP), Galvanic Stair-Step (ISTEP)
Chrono Methods
Chronopotentiometry (CP), Chronoamperometry (CA), Chronocoulometry (CC)
Impedance
EIS vs Frequency (IMP), EIS vs Time (IMPT), EIS vs Potential (IMPE)
Corrosion testing
Electrochemical Noise (EN), Zero Resistance Ammeter (ZRA), Electrochemical Potentiokinetic Reactivation(EPR)
Battery testing
Battery charge and discharge, Galvanostatic charge and discharge
Extensions
Data Logger, Bulk electrolysis with Coulometry


 