Specifications |
Support 2-, 3- or 4-electrode system |
Potential control range: Primary Channel: ±10V
Second Channel: ±10V |
Current control range: ±2A |
Potential control accuracy: 0.1%×full range±1mV |
Current control accuracy: 0.1%×full range |
Potential resolution: 10μV (>100Hz),3μV (<10Hz) |
Current sensitivity:1pA |
Rise time: <1μS (<10mA), <10μS (<2A) |
Reference electrode input impedance:1012Ω||20pF |
Current range: 2nA~2A, 10 ranges |
Compliance voltage: ±21V |
Maximum current output: 2A |
CV and LSV scan rate : 0.001mV~10,000V/s |
CA and CC pulse width: 0.0001~65,000s |
Current increment during scan: 1mA@1A/ms |
Potential increment during scan: 0.076mV@1V/ms |
SWV frequency: 0.001~100 kHz |
DPV and NPV pulse width: 0.0001~1000s |
AD data acquisition:16bit@1 MHz,20bit@1 kHz |
DA Resolution:16bit, setup time:1μs |
Minimum potential increment in CV: 0.075mV |
IMP frequency: 10μHz~1MHz |
Low-pass filters: Covering 8-decade |
Potential and current range: Automatic |
PC and O/S requirement |
Interface: USB 2.0 |
Operating System: |
Windows 2000/NT/XP/ win7/win8/win10 |
Weight / Measurements |
Net weight: 6.5kg Gross weight: 10kg |
Instrument size: 36.5 x 30.5 x16cm; After package: 52.5 x 37.5x 22.5cm |
Electrochemical Impedance Spectroscopy(EIS) |
Signal generator |
Frequency range:10μHz~1MHz |
AC amplitude:1mV~2500mV |
DC Bias:-10~+10V |
Output impedance:50Ω |
Waveform: sine wave, triangular wave and square wave |
Wave distortion:<1% |
Scanning mode: logarithmic/linear, increase/decrease |
Signal analyzer |
Integral time:minimum:10ms or the longest time of a cycle |
Maximum:106 cycles or 105s |
Measurement delay:0~105s |
DC offset compensation |
Potential automatic compensation range:-10V~+10V |
Current compensation range:-1A~+1A |
Bandwidth: 8-decade frequency range, automatic and manual setting |